INDEX
- PRESSURE MEASUREMENT Q&A
- LEVEL MEASUREMENT Q&A
- FLOW MEASUREMENT Q&A
- TEMPERATURE MEASUREMENT Q&A
- FIRE AND GAS SYSREM Q&A
PRESSURE CONVERSIONS :
1psi =
|
27.74 " H2O
|
1 Kg/cm2 =
|
14.223 psi
|
1 Bar =
|
14.504 psi
|
1 Kpa =
|
0.145 psi
|
1 Kg/cm2 =
|
10.000mm of H20
|
1 Bar =
|
1.0197 Kg/cm2
|
1 Kg/cm2 =
|
0.98 Bar
|
1 Torr =
|
1 mm of Hg.
|
Pressure: It is defined as Force per unit Area. P = F/A
Units :
bar, Pascal, kg / cm2, lb / in2.
Que. : What is
absolute pressure ?
Ans. : Absolute
pressure is the total pressure present in the system
Abs. pressure = Gauge pressure + Atm. pressure.
Que. : What is
absolute zero pressure ?
Ans. : Absolute
zero = 760 mm Hg Vacuum.
Que. : What is
the maximum Vacuum ?
Ans. : The
maximum Vacuum = 760 mm Hg.
Que. : What is
Vacuum ?
Ans. : Any
pressure below atmospheric pressure is vacuum.
Atm = 760 mm 0 Vacuum Zero
Gauge
_______________________________
'
'
' _______
Max. Vacuum 760 mm HG
Zero
Absolute.
Que. : What
are the primary elements for measuring pressure ?
Ans. : The
primary elements used for measuring pressure are :
1. Bourdon tube.
2. Diaphragm.
3. Capsule.
4. Bellows.
5. Pressure Springs.
The above are known as elastic deformation pressure
elements.
Type of Bourdon tubes.
1. ' C ' type.
2. Spiral.
3. Helix.
Diaphragm : The diaphragm is best suited for low pressure
measurement.
Capsules : Two circular diaphragms are welded together to form
a pressure capsule.
Material Used : phosphor bronze, Ni-spanc stainless
steel.
Bellows : Bellows is a one - piece, collapsible, seamless
metallic unit with deep folds formed from very tin walled tubing.
Material used : Brass, phosphor bronze, stainless
steel. Used for high pressure.
PRESSURE SPRINGS : Pressure springs of helical or spiral shape
used for measuring high pressures.
Que. : What is
the principle of a pressure gauge ?
Ans. : Pressure
works on Hooks law.
Principle : "Measuring the stores in an elastic
medium"
Que. : Draw
and explain a pressure gauge ? What is the used of a Hair spring ?
Ans. : The
parts of a pressure gauge are :
1. 'C' type Bourdon tube.
2. Connecting link.
3. Sector gear.
4. Pinion gear.
5. Hair Spring.
6. Pointer.
7. Dial.
Uses of Hair Spring : Hair spring serves two
purposes name
1. To eliminate any play into linkages.
2. It serves as a controlling torque.
Que. : How
will you calibrate an absolute pressure transmitter vacuum manometer Range 0-400mm abs. ?
Ans. : The
procedure for calibration is as follows :
1. Connect air supply to the transmitter.
2. Connect a test gauge of 0-1.4 Kg/cm2 to the output.
3. Connect Vacuum pump with tee off to the
manometer.
4. Apply 760 mm Vacuum ( or nearest ) and adjust
zero.
5. Apply 360 mm Vacuum adjust span. ( 760 - 360 =
400 mm abs. )
Que. : You are
given a mercury manometer range 0 -760 mm ? A vacuum gauge reads 60 mm vacuum.
The test manometer reads 50 vacuum ? Which of the two in correct.
Ans. : The
transmitter is correct because 760 - 50 = 710 mm abs.
1.what are the application advantages of a dead weight tester and a
gauge comparator?
Dead weight tester: Generally it is used for calibrating a test gauge.
It works on the hydraulic principle, where a test gauge is compared with the
standard dead weights.
Calibration procedure is laborious.
Gauge comparator: It is used for calibrating a pressure gauge against a
test gauge. Calibration procedure is simple and quicker.
2.How would you select a pressure gauge for a process?
While selecting a pressure gauge for a process, consider the following
points:
_ Characteristic of the process
(corrosive or non-corrosive)
_ Operating pressure
_ Maximum process pressure
_ The gauge range should be a
minimum of twice the process operating pressure.
3. Write the types of Bourdon tubes? Explain the purpose of different
Bourdon tubes.
The following are the commonly used Bourdon tubes used in industries:
·
Spiral
–Low range 10 –100kpa
·
C type
–Medium range 100 – 5000kpa
·
Helical
– High range 5000 – 20000kpa
(refer to the manufacturers manuals for
correct parameters)
4.What is function of a hair –spring in a pressure gauge?
Hair – spring in a pressure gage eliminates the Hysterisis
(backlash/angularity) error caused in the quadrant (gear and pinion mechanism)
and the weight of the pointer.
5.Name a few pressure switch manufacturing companies.
Barksdale, United electric, CCS…etc.
6.What is the output of a pressure switch?
Output of an electric switch is a contact- open or close.
Out put of a pneumatic switch is a port operation- open or close, vent
7.How would you select a pressure switch for 500kpa Hi operatin?
The following points are to be considered wile selecting a pressure
switch for any process operation:
·
The
process operating pressure
·
The
maximum process pressure
·
The
process pipeline vibration
·
The
maximum working pressure of the pressure switch should be times greater then
the maximum operating pressure
·
The
micro switch contact rating
·
Process
connection
·
Local
or remote mounting
·
…etc.
8.Expain the contact selection on high and low pressure alarm switches?
And explain why?
On a high pressure switch, the wiring is terminated on the “common” and
the “normally close” contact terminals.
On a low pressure switch, the wiring is terminated on the “common” and
the “normally open” contact terminals.
This type of contact termination is done to achieve a close contact from
the switch during a normal (healthy) process condition which is a fail safe
method.
9.What is to be done, if a transmitter gives a maximum output , where
the transmitter range (jumper) is already in maximum selection?
Replace the capsule (sensor) for a higher range in the transmitter.
10.Explain how to carryout a field zero check on a pressure transmitter?
·
If the
transmitter signal is used for controlling, then put the controller on manual.
·
Connect
a digital multimeter in the current range in series with the transmitter
output.
·
Isolate
the process valve
·
Isolate
the instrument isolation valve.
·
Open
the equalising valve
·
Open
the bleed valve and de-pressurise the transmitter.
·
Check
for 4.00mA output signal, if not, adjust the zero screw.
11.What-test equipment is required to calibrate a pressure transmitter
in the field?
If it is a low range transmitter, then precision pneumatic test
equipment like “wallace and tiernier” and pneumatic pump (or air supply) may be
used.
If it is a high range transmitter, then a hydraulic pump and a test
gauge is to be used.
12.What are pressure measuring/feeding test equipment?
The following are a few commonly used test equipment to measure and feed
pressure: Tradinco, Wallance & tiernier, Ralston, Gauge comparator, Dead
weight tester…etc.
Dead weight tester is used for calibrating a
test gauge. It works on the hydraulic principle, where as test gauge is
compared with the standard dead weights n dead weight tester.
Gauge comparator: It is used for calibrating
a pressure gauge against a test gauge. Calibration procedure is simple and
quicker.
2.How would you select a
pressure gauge for a process?
While selecting a pressure gauge for a
process, consider the following points:
#Characteristic of the process (corrosive or
non-corrosive)
#Operating pressure
#Maximum process pressure
#The gauge range should be a minimum of twice
the process operating pressure.
3. Write the types of Bourdon
tubes? Explain the purpose of different Bourdon tubes.
The following are the commonly used Bourdon
tubes used in industries:
#Spiral –Low range 10 –100kpa
#C type –Medium range 100 – 5000kpa
#Helical – High range 5000 – 20000kpa
4.What is function of a
hair–spring in a pressure gauge?
Hair–spring in a pressure gauge eliminates
the Hysterisis (backlash/angularity) error caused in the quadrant (gear and
pinion mechanism) and the weight of the pointer.
5.Name a few pressure switch manufacturing
companies.
Barksdale, United electricals, don foss,
CCS…etc.
6.What is the output of a
pressure switch?
Output of an electric switch is a contact-
open or close.
Out put of a pneumatic switch is a port operation- open or close, vent
7.How would you select a pressure switch for
500kpa Hi operatin?
The following points are to be considered
wile selecting a pressure switch for any process operation:
#The process operating pressure
#The maximum process pressure
#The process pipeline vibration
#The maximum working pressure of the pressure
switch should be 2 times greater
then the maximum operating pressure
#The micro switch contact rating
#Process connection
#Local or remote mounting
8.Expain the contact selection
on high and low pressure alarm switches? And explain why?
On a high pressure switch, the wiring is
terminated on the “common” and the “normally close” contact terminals.
On a low pressure switch, the wiring is
terminated on the “common” and the “normally open” contact terminals.
This type of contact termination is done to
achieve a close contact from the switch during a normal (healthy) process
condition which is a fail safe method.
9.What is to be done, if a
transmitter gives a maximum output , where the transmitter range (jumper) is
already in maximum selection?
Replace the capsule (sensor) for a higher
range in the transmitter.
10.Explain how to carryout a field zero check
on a pressure transmitter?
1.If the transmitter signal is used for
controlling, then put the controller on
manual.
2.Connect a digital multimeter in the current
range in series with the transmitter output.
3.Isolate the process valve
4.Isolate the instrument isolation valve.
5.Open the equalising valve
6.Open the bleed valve and de-pressurise the
transmitter.
7.Check for 4.00mA output signal, if not,
adjust the zero screw.
11.What-test equipment is
required to calibrate a pressure transmitter in the field?
If it is a low range transmitter, then
precision pneumatic test equipment like “wallace and tiernier” and pneumatic
pump (or air supply) may be used.
If it is a high range transmitter, then a
hydraulic pump and a test gauge is to be used.
12.What are pressure
measuring/feeding test equipment?
The following are a
few commonly used test equipment to measure and feed pressure: Tradinco,
Wallance & tiernier, Ralston, Gauge comparator, Dead weight tester…etc.
2. L E V E L
Level: The height of the water column, liquid and powder
etc., at the desired measurement of height between minimum level points to
maximum level point is called level. The measurement principle is, head
pressure method.
Units: Meters, mm, cm, percentage.
Que. : Explain
the different methods of level measurement in brief ?
Ans. : There
are two ways of measuring level :
( 1
) Direct & ( 2 ) Indirect.
( 1 ) DIRECT LEVEL MEASUREMENT :
1. BOB AND TOP :
2. SIGHT GLASS : This consists of a graduated
glass tube mounted on the side of the vessel. As the level of the liquid in the
vessel change, so does the level of the liquid in the glass tube.
( 2 ) INDIRECT LIQUID LEVELMEASUREMENT :
1. PRESSURE GAUGE : This is the simplest
method, for pressure gauge is located at the zero level
of the liquid in the vessel. Any rise in level causes an increase of pressure which can be measured by a gauge.
2. PURGE SYSTEM : In this method a pipe is installed vertically
with the open and at zero level. The other end of the pipe is connected to a regulated air r supply and to a pressure gauge. To make a level measurement
the air supply is adjusted so that pressure is slightly higher than the
pressure due to height of the liquid.
This is accomplished by regulating the air pressure until bubbles cab be seen
slowly leaving the open end of the pipe.
The method above are suitable for open tank
applications. when a liquid is in a pressure vessel, the liquid column pressure
can't be used unless the vessel pressure is balanced out. This is done through
the use of different pressure meters.
DIFF.PRESSURE METHOD : Connection are made at the
vessel top and bottom, and to the two columns of the D.P. meter. The top
connection is made to the L.P. column of the transmitter and the bottom to H.P.
column of the transmitter. The difference in pressure in the vessel is balanced
out, since it is fed to both the column of the meter. The difference in
pressure deducted by the meter will be due only to the changing, level of the
liquid.
DISPLACER TYPE LEVEL MEASUREMENT :
The leveltrol is one of the most common instruments
used measuring level in closed tanks. This instrument works of Archimedes
principle. The displacer in immersed in the liquid due to which there is loss
of weight depending on the specified gravity of the liquid. This displacer
hangs freely on a knife transmitted to the pneumatic or electronic counterpart
at the other end.
Que. : Explain
how you will measure level with a different pressure transmitter.
Ans. : The
bottom connection of the vessel is connected to high pressure side of the
transmitter.
Different Pressure
= H X D
This difference pressure is applied to H.P. side of
the transmitted and calibrated.
Que. : How is
D.P. transmitter applied to a close tank ?
Ans. : In
close tank the bottom of the tank is connected to the high pressure side of the
transmitter and top of the tank in connected to L.P. side of the transmitter.
In this way the vessel pressure is balanced.
Que. : How is
D.P. transmitter applied to an open tank ?
Ans. : On an
open tank level measurement the L.P. side is vented to atmosphere. Whatever
pressure acts is on the H.P. side which is a measure of level.
Que. : What is
purge level system ?
Ans. : This
method is also known as bubbler method of level measurement. A pipe is
installed vertically with its open end at the zero level. The other end of the
pipe is connected to a regulated air supply and to a pressure gauge or to ^P
transmitter. To make a level measurement the air supply is adjusted so that
pressure is slightly higher than the pressure due to the height of the liquid.
This is accomplished by regulating the air pressure until bubbles can be seen
slowly leaving the open end of the pipe. The gage then measures the air
pressure needed to over come the pressure of the liquid.
/\ P = H X D
USE : On for corrosive liquids where the transmitter
cannot be directly connected to process eg... Acids, Some organic liquids.
Que. : Explain
the working of a leveltrol.
Ans. : The
leveltrol is used for measuring level 0 liquids in a closed vessel.
PRINCIPLE. : It works on Archimedes principle
"The loss in weight of a body immersed in a liquid is equal to amount of
liquid displaced by the body". The leveltrol basically consists of the
following :
DISPLACER : It is consists of a cylindrical shape
pipe sealed and filled inside with sand or some weight. The purpose of this is
to convert change in level to primary motion. The variation in buoyancy
resulting from a change in liquid level varies the net weight of the displacer
increasing or decreasing the load on the torque arm. This change is directly
proportional to change in level and specific gravity of the liquid.
RELAY : Amplifies pressure variations at the
nozzles.
REVERSING ARC : It is used for the following
purposes.
1. Motion take of from Torque tube.
2. Means of reverse control action.
3. Adjustment for specific gravity.
PROPERTIONAL UNIT. : Converts primary motion to a
proportional output air pressure.
CONTROL SETTING UNIT : Provides a motions of varying
the set point.
Que. : Explain
the working an electronic leveltrol.
Ans. : The
variation in buoyancy resulting from a change in liquid level, varies the net
weight of the displacer increasing or decreasing the load on the torque arm.
This change is directly proportional to the change in level and specific
gravity of the liquid. The resulting torque tube movement varies the angular
motion of the rotor in the RVDT (Rotary Variable Differential. Transformer)
providing a voltage change proportional to the rotor displacement, which is
converted and amplified to a D.C. current.
Que. : How
will you reverse an action of the leveltrol.?
Ans. : The
reversing are serves as motion take off arm from the torque tube. It is
provided with a slot on each side of the center so that link can be connected
either for reverse or direct action.
Que. : What is
interface level ? How do you calculate it ?
Ans. : When a
vessel is filled with two liquids of two different specific gravities the level
measurement refers to as interface level.
DP = H ( D -
d )
On a level set the difference of two specific
gravities.
Que. : How
will you calibrate a leveltrol in the field ?
Ans. : 1.
First close both the primary isolation valves and drain the liquid inside the
chamber.
2. Adjust the zero to get 0% output.
3. Connect a transparent PVC tube to the drain point
as shown in hook up.
4. Fill it to the center of the top flange.
5. Adjust the specific gravity or span
adjustment ( Electronic Level ).
6. Fill it up to 50 %, check linearity.
5. To check linearity apply average of the two
weights.
Que. : What
will happen if the displacer has fallen down while in line ?
Ans. : The
output will be minimum. ( wrong )
The
output will be maximum (100 %) true.
Que. : What
will happen if the displacer has a hole in it while in line ?
Ans. : The
output will be maximum.
Que. : What is
the used of Suppression and elevation ?
Ans. : Suppression
and elevation are used on Level applications where (1) transmitters are not
mounted on some level (2) Wet leg. i.e. condensable vapors are present.
Que. : What
are the limitations of leveltrol ?
Ans. : The
limitations of a level control that it cannot be used for lengths more than
72".
Que. : How
will you commission D.P. transmitter in field in pressurized vessel.
Ans. : 1.
Close both the isolation valves, Vent the H.P. side.
2.
Fill it with the sealing liquid.
3.
Open the L.P. side vent valve.
4.
Adjust zero with suppression spring.
5.
Close the L.P. side vent valve.
6.
Open both the isolation valves.
Que. : How
will you check zero of a level D.P. transmitter while is line ?
Ans. : 1.
Close both the isolation valves.
2.
Open the vent valve on L.P. leg and H.P. leg drain.
3.
Check and adjust zero if necessary.
Que. : Explain
the working of an Enraf level gauge ?
Ans. : The
Enraf level precise level gauges are based on servo powered null balance
technique. A displacer serves as continuos level sensing element.
Principle :
A displacer A with a relative density higher than
that of a product to be measured, is suspended from a stainless steal wire B
tat is attached to a measuring drum. A two phase servo motor controlled by a
capacitive balance system winds unwinds the measuring wire until the tension on
the weight springs is in balance with the wt. of the displace part immersed in
the liquid. The sensing system in principle measures the two capacitance formed
by the moving central sensing rod E provided with two capacitor plates and the
si plates. In balance position the capacitance are of equip value. A level
variation will cause a difference in buoyancy of the displacer. The center
sensing rod will move in to direction of one of the side capacitor plates. This
causes difference in value of this capacitance. By an electrolyte rotation of
the servo motors the can driven transmitter continuously change the voltage
pattern to remote indicate of which the
receiver motor drives a counter indicating low variation.
h= height of the
liquid column
2.What are the simple methods for measuring level?
Gauge glass, Pressure Gauge – scale graduated in % of level, Rope and
weight , Dip Tape…etc
3.How to convert a pressure gauge
into a level gauge?
Calculate the static head in kpa using the formula “pgh”. Select a
pressure gauge and calibrate it for the calculated static head. Graduate the
pressure gauge scale in terms of %of level.
4.What is the density of crude oil?
The density of crude oil is approximately 0.8.
5.What is an “interface dip tape”? Where it used?
Interface dip tape is an instrument used for measuring the total and the
interface liquid level in vessel.
For example: It is used in oil storage tanks to measure the total liquid
level and water level.
6.Explain how a Leveltrol works? Name the parts of a pneumatic
Leveltrol.
A Leveltrol in and instrument used for measuring the liquid level
between two known points. The Leveltrol works on the buoyancy principal.
Leveltrol has a float, which submerses proportionately with liquid level raise
in the float chamber. The amount of submersion of the displacer depends on the
liquid density, which produces a torque. The amount of torque produced in
measured in terms of % of level.
Parts of Leveltrol:
Float chamber, Float, Torque lever, Knife edge, Feedback Bellows, Air
Relay, Restriction, Flapper, Nozzle, Feedback link, Density range, Action
change lever, HP and LP flange,
7.What is a static head level transmitter?
A static head level transmitter is used for measuring the total liquid
level in the tank.
8.What are the application limits of a Leveltrol and a static head level
transmitter?
A Leveltrol can measure only for a short and fixed height of level. A
Leveltrol has a high gain output. In can be used for liquid level as well for
the interface level measurement.
Static head transmitter is used for measuring the total liquid level in
the tank. The transmitter output is linear.
9.What is the operating principle of a Magnetrol level switch?
The Magnetrol level switch works on the principle of Buoyancy force.
10.What are the points to consider while selecting a level switch for a
particular process?
While selecting a level switch, the following points are to be
considered:
·
The
characteristic of process (corrosive or non – corrosive)
·
The
process pressure
·
The
liquid density
·
The flange
ratings
·
Proof
pressure of the switch
·
The
micro switch contact rating
11.What type of level measuring instrument is suitable for closed tank?
If the level measurement is required for a fixed and small height, a
Leveltrol is more suitable.
If it is for a total height, then either a static head or a differential
pressure transmitter with its HP leg connected to the bottom of the liquid
level and the LP leg connected to the top of the tank to gas phase.
12.What is Zero suppression and Zero elevation in level measurement?
Zero suppression:
when a static head transmitter is installed below the zero liquid level, the
transmitter gets a +ve error in the level measurement. This error is
corrected by a zero suppression kit.
Zero elevation: when
a static head transmitter is installed above the zero liquid level, the
transmitter gets a –ve error in the level measurement. The error is corrected
by a zero elevation kit.
13.What is an interface in level measurement?
An interface is the separation point between the two de-missible liquids
levels In a vessel. This condition arises when the liquid does not mix due to
its chemical composition and difference in their density.
14.How to calibrate a leveltrol for an interface level measurement?
Fill the leveltrol chamber 100% with the lower density liquid and adjust
its zero for 4.00mA output.
Drain the liquid and fill the leveltrol chamber 100% with the higher
density liquid and adjust its span for 20.00 mA output.
The transmitter on line measures the percentage of higher density liquid
in the lower density liquid at a known height.
15.How to calibrate a static head level transmitter for an interface
measurement?
Static head level transmitter is not commonly used for measuring the
interface level. But the following procedure may be used for calibrating it to
measure the interface level in a tank.
Fill the vessel 100% with the lower density liquid and adjusts its zero
for 4.00 mA output. Drain the liquid and fill with the vessel 100% with the
higher density liquid and adjust its span for 20.00mA output. 100% transmitter
level is to be continuously maintained.
The transmitter on line measures the percentage of higher density liquid
in the lower density liquid in a know height.
16.Why is a leveltrol more suitable tan a static head level transmitter
on a separator?
Leveltrol is more accurate in measuring the small height of liquid
level. The process pressure change does not affect the leveltrols performance.
Leveltrol offers a high gain output verses the change in the liquid level.
17.Why is static head level transmitter more suitable than a leveltrol
on a surge tank?
A leveltrol has a limitation in measuring the maximum height of liquid.
Considering huge of the surge tank, a static head level transmitter is more
suitable.
18. Why is a gap controller often used in controlling a level in a high
pressure vessel?
Gap controllers output changes from minimum to maximum and visa versa
when the process measurement deviates the setpoint by the set gap. Hence the final
control element i.e. control valve acts like an on-off valve. This helps in
minimizing the trim erosion due to a high DP across the control valve while
partially open.
19. How to use a DP (differential pressure) transmitter for level
measurement?
A differential pressure transmitter gives a linear output for the
differential pressure measured across its HP and LP chambers.
Connect the transmitter’s HP leg to the bottom of the tank and its LP
leg to the top of the tank. The transmitter will read the liquid level
accurately irrespective of the change in process pressure above the liquid
surface.
Indirect level measurement:
(A) Pressure gauge:
This
is the simplest method, for pressure gauge is located at the zero level of the
liquid in the vessel. Any rise in level causes an increase of pressure which
can be measured by a gauge.
(b) Purge system:
In
this method a pipe is installed vertically with the open and at zero level. The
other end of the pipe is connected to a regulated air r supply and to a pressure
gauge. To make a level measurement the air supply is adjusted so that pressure
is slightly higher than the pressure due to height of the liquid. This is
accomplished by regulating the air pressure until bubbles cab be seen slowly
leaving the open end of the pipe.
The air pressure to the bubbler pipe
is minutely in Excess
of the liquid pressure in the vessel, so that Air
pressure indicated is a measure of the level in The
tank.
The
method above is suitable for open tank applications. When a liquid is in a
pressure vessel, the liquid column pressure can't be used unless the vessel
pressure is balanced out. This is done through the use of different pressure
meters.
(c)
Differential pressure meter:
Connections
are made at the vessel top and bottom, and to the two columns of the D.P.
meter. The top connection is made to the L.P. column of the transmitter and the
bottom to H.P. column of the transmitter. The difference in pressure in the
vessel is balanced out, since it is fed to both the column of the meter. The
difference in pressure deducted by the meter will be due only to the changing,
level of the liquid.
(d) Displacer type
level measurement:
The
leveltrol is one of the most common instruments used measuring level in closed
tanks. This instrument works of Archimedes principle. The displacer in immersed
in the liquid due to which there is loss of weight depending on the specified
gravity of the liquid. This displacer hangs freely on a knife transmitted to
the pneumatic or electronic counterpart at the other end.
30.
Explain how you will measure level with a different pressure transmitter.
The
bottom connection of the vessel is connected to high pressure side of the
transmitter.
Different
Pressure = H X D
This
difference pressure is applied to H.P. side of the transmitted and calibrated.
31.
How is D.P. transmitter applied to a close tank?
In
close tank the bottom of the tank is connected to the high pressure side of the
transmitter and top of the tank in connected to L.P. side of the transmitter.
In this way the vessel pressure is balanced.
32.
How is D.P. transmitter applied to an open tank?
On
an open tank level measurement the L.P. side is vented to atmosphere. Whatever
pressure acts is on the H.P. side which is a measure of level.
SPAN = (X) (Sp. Graf)
ZERO SUPPRESSION = (Y) (Sp.Grav)
ZERO SUPPRESSION = (Y) (Sp.Grav)
33.
How is D.P transmitter applied to a close tank & open tank with Dry leg?
Span = (X) (GL)
HW
at minimum level = ( Z ) ( GS ) + ( Y
) ( GL )
HW
at maximum level = ( Z ) ( GS
) +
( X + Y ) ( GL )
Where:
GL =
Specific gravity of tank liquid.
GS =
Specific gravity of seal liquid.
HW =
Equivalent head of water.
X, Y & Z are
shown in fig (1.1)
Example:
Open tank with
X = 300 inches
Y = 50 inches
Z = 10 inches
GL = 0.8
GS = 0.9
Span = (300) (0.8) = 240 inches
HW at minimum level = ( 10 ) ( 0.9 ) + ( 50
) ( 0.8 ) = 49 inches
HW at maximum level = (10 ) ( 0.9 )
+ ( 50 + 300 ) ( 0.8 ) = 289
inches
Calibrated
range = 49 to 289 inches head of water
2.3 DEFINITION OF INSTRUMENT RANGE AND SPAN
Legend:
Range Defined
in IEC 60902 as follows: The region of values between the lower and upper
limits of the quantity under consideration.
It is expressed by stating the lower and upper limits (e.g. minus 1 to 10 bars (ga)).
It is expressed by stating the lower and upper limits (e.g. minus 1 to 10 bars (ga)).
Span Defined
in IEC 60902 as follows: The algebraic difference between the upper and
lower limit values of a given range.
It is expressed as a figure and unit of measurement (e.g. 8 bars).
It is expressed as a figure and unit of measurement (e.g. 8 bars).
LRL Lower
Range Limit; the lowest quantity that a device is designed to measure.
URL Upper
Range Limit; the highest quantity that a device is designed to measure.
Instrument
range the
region in which the instrument is designed to operate. It is a physical
capability of the device. The region limits are expressed by stating the LRL
and URL.
Instrument
minimum span the minimum
distance between the URV and LRV for which the instrument is designed. It is a
physical limitation of the device.
Instrument
maximum span The maximum
distance between the URV and LRV for which the instrument is designed. It is a
physical limitation of the device.
LRV Lower
Range Value; the lowest quantity that a device is adjusted to measure.
URV Upper
Range Value; the highest quantity that a device is adjusted to measure.
Adjusted
range The
measurement region. It is expressed by stating the LRV and URV.
Adjusted
span The
distance between the URV and LRV.
Example: A differential
pressure transmitter is used to measure the level in a vessel, using a wet
reference leg. 0% level corresponds with a differential pressure of -800 mbar
and 100% level with - 100 mbar. The catalogue of the selected transmitter lists
-1800/+1800 mbar for LRL/URL respectively and span limits of 300 to 1800 mbar,
so:
Instrument
range
|
=
|
-1800 to
+1800 mbar
|
Instrument
minimum/maximum span
|
=
|
300 / 1800
mbar respectively;
|
LRV / URV
|
=
|
-800 /
-100 mbar respectively;
|
Adjusted
range
|
=
|
-800 to
-100 mbar.
|
Adjusted span
|
=
|
700 mbar.
|
NOTE: ‘Adjusted’ range
and ‘adjusted’ span are frequently referred to as ‘calibrated’ range and
‘calibrated’ span. This term is however only correct, if a calibration facility
is used to set the LRV and URV. For ‘intelligent’ measuring devices, the
supplier is usually calibrating the device at the LRL/URL and the user is
setting the required LRV and URV by remote communication.
2.4 SELECTION OF RANGES
The accuracy (2.8) and adjusted range of an instrument
should be selected to cover the operating window (2.1), which includes
applicable abnormal operation and alternative operating modes. Unless otherwise
stated, the normal design value should lie between 50% and 75% of the adjusted
range
NOTE: For certain applications it might not be
possible to combine all normal and abnormal operating conditions in one
measurement at the required accuracy. In such cases, a case-by-case analysis
should disclose whether additional instruments are required or the accuracy
requirements and/or operating window may be relaxed. It might be acceptable to
present measured values during some of the abnormal operating cases at a lower
accuracy or it might be justifiable for the measurement not to produce a
sensible signal under some of the abnormal process conditions during start-up,
commissioning, regeneration, emergency conditions and the like.
IPF transmitters should have the same instrument range,
adjusted range and accuracy as corresponding process transmitters in order to
facilitate measurement comparison. For details and exceptions, see
DEP 32.80.10.10‑Gen. Trip settings should lie between 10% and 90% of the
adjusted range.
The LRV should be selected so that the displayed result
represents the zero or sub-zero value of the process variable (e.g. 0-150
tons/day, 0-100% level, 0-10 bar (ga), -1/+3 bar (ga), 0‑500 °C,
-50/+50 °C etc.). Elevated zero’s (100-300 tons/day, 100-200 °C)
should be avoided.
Close tank with wet leg:
Span = (X) (GL)
HW at minimum level = (
Y ) ( GL ) – ( d )( GS )
HW at maximum level = ( X
+ Y ) ( GL ) – ( d ) ( GS )
Where: GL = Specific gravity of tank liquid
GS =
Specific gravity of tank liquid
HW = Equivalent head of water
X, Y and Z are shown in fig.
Example :
X = 300
inches
Y = 50
inches
d = 500
inches
GL = 0.8
GS = 0.9
Span = ( 300 ) ( 0.8 ) = 240
inches
HW
minimum level = ( 50 ) ( 0.8 ) - (
500 ) ( 0.9 ) = - 410 inches
HW
maximum level = ( 300 + 50 )
( 0.8 ) – ( 500 ) ( 0.9 ) = - 170 inches
Calibrated
range =
- 410 to –170 inches head of water.
(
minus sings indicate that the higher pressure is applied to the low pressure
side of the transmitter )
34.
What is purge level system?
This
method is also known as bubbler method of level measurement. A pipe is installed
vertically with its open end at the zero level. The other end of the pipe is
connected to a regulated air supply and to a pressure gauge or to ^P
transmitter. To make a level measurement the air supply is adjusted so that
pressure is slightly higher than the pressure due to the height of the liquid.
This is accomplished by regulating the air pressure until bubbles can be seen
slowly leaving the open end of the pipe. The gage then measures the air
pressure needed to over come the pressure of the liquid.
/\ P = H X D
USE:
On for corrosive liquids where the
transmitter cannot be directly connected to process eg... Acids, Some organic
liquids.
35.
Explain the working of a leveltrol.
The
leveltrol is used for measuring level of liquids in a closed vessel.
1. PRINCIPLE. : It works on Archimedes principle
"The loss in weight of a body immersed in a liquid is equal to amount of
liquid displaced by the body". The leveltrol basically consists of the
following :
2. DISPLACER: It is consists of a cylindrical
shape pipe sealed and filled inside with sand or some weight. The purpose of
this is to convert change in level to primary motion. The variation in buoyancy
resulting from a change in liquid level varies the net weight of the displacer
increasing or decreasing the load on the torque arm. This change is directly
proportional to change in level and specific gravity of the liquid.
3. RELAY: Amplifies pressure variations at the
nozzles.
4. REVERSING
ARC: It is used for the following purposes.
·
Motions take of
from Torque tube.
·
Means of reverse
control action.
·
Adjustment for specific gravity.
5. PROPERTIONAL UNIT. : Converts primary motion to
a proportional output air pressure.
6. CONTROL SETTING UNIT: Provides motions of
varying the set point.
36.
Explain the working an electronic leveltrol.
The
variation in buoyancy resulting from a change in liquid level, varies the net
weight of the displacer increasing or decreasing the load on the torque arm.
This change is directly proportional to the change in level and specific
gravity of the liquid. The resulting torque tube movement varies the angular
motion of the rotor in the RVDT (Rotary Variable Differential. Transformer)
providing a voltage change proportional to the rotor displacement, which is
converted and amplified to a D.C. current.
37.
How will you reverse an action of the leveltrol?
The
reversing are serves as motion take off arm from the torque tube. It is
provided with a slot on each side of the center so that link can be connected
either for reverse or direct action.
38.
What is interface level? How do you calculate it?
When
a vessel is filled with two liquids of two different specific gravities the
level measurement refers to as interface level.
DP = H (D - d)
On
a level set the difference of two specific gravities.
39.
How will you calibrate a leveltrol in the field?
Calculation
# 1 If the calibrating liquid is water:
Process Liquid Density / 1* Displacer height = mm of H2O.
Calculation
# 2 If the calibrating liquid is other liquid:
Process Liquid Density / Calibrating liquid density * Displacer height =
mm of H2O. (Calibrating liquid height in mm).
1.
First close both the primary isolation valves and drain the liquid inside the
chamber.
2.
Adjust the zero to get 0% output.
3.
Connect a transparent PVC tube to the drain point as shown in hook up.
4.
Fill it to the center of the top flange.
5.
Adjust the specific gravity or span adjustment (Electronic Level).
6.
Fill it up to 50 %, check linearity.
40.
How will you calibrate on interface level control. ?
On
an interface leveltrol there are two liquid of two different specific
gravities.
1.
The level will be zero when it is full of lighter liquid.
Zero % level = H X d.
H = Displacer length
d = Specific gravity of lighter liquid.
2
The level will be 100 % when it is full of heavier liquid.
100 % level = H X D.
D = Specific
gravity of heavier liquid.
Calibration
with water:
1.
Fill H X d level with water adjust zero.
2.
Fill H X D level with water adjust Sp. gravity or span.
3.
Check linearity.
Y= L* (DH X + (1- X ) * dl) ------- ------
100 100
|
Displacer length: L in mm
Density Low
: dl in Kg / 1
Density High
: DH in Kg / 1
Alarm / Trip point %: X
HH Trip Point:
Y in mm
41.
How will you apply wt. lest calibration to a leveltrol.
Wt. test calibration method:
- Remove the displacer
from the torque arm.
- Apply equivalent
weight on the torque arm that is equal to the wt. of the displacer. Adjust
zero % output.
- For Span : V = πr2h
Loss in weight = Wt. of float - wt. of the float
immersed in liquid
Loss in weight = [ wt. of float - Vol. x d ]
Span wt. = (wt. of float - Loss
in wt.)
r = radius of the displacer.
h = ht. of displacer.
- Apply equivalent wt.
equal to the (Wt. of float - Loss in weight). Adjust Span to get 100 % out
put.
- To check linearity
apply average of the two weights.
42.
What will happen if the displacer has fallen down while in line ?
The
output will be maximum.
43.
What will happen if the displacer has a hole in it while in line?
The
output will be minimum.
44.
What is the used of Suppression and elevation?
Suppression
and elevation are used on Level applications where (1) transmitters are not
mounted on some level (2) Wet leg. I.e. condensable vapors are present.
45.
What are the limitations of leveltrol?
The
limitations of a level control that it cannot be used for lengths more than 72
inches.
46.
How will you commission D.P. transmitter in field in pressurized vessel.
1. Close both the isolation valves, Vent the H.P. side.
2. Fill it with the sealing liquid.
3. Open the L.P. side vent valve.
4. Adjust zero with suppression spring.
5. Close
the L.P. side vent valve.
6. Open both the isolation valves.
47.
How will you check zero of a level D.P. transmitter while is line?
1. Close both the isolation valves.
2. Open the vent valve on L.P. leg and H.P. leg drain.
3. Check and adjust zero if necessary.
48.
Explain the working of an Enraf level gauge?
The
Enraf precise level gauge are based on servo powered null-balance technique. A
displacer serves as a continuous level sensing element.
Principle:
A
displacer with a relative density higher than that of the product to be
measured, is suspended from a stainless steel wire B, that is attached to a
measuring drum. A two phase servo meter controlled by a capacitive balance
system winds or unwinds the measuring wire until the tension in the weighing
springs is in balance with the weight of the displacer partly immersed in the
liquid. The sensing system in principle measures the two capacitance formed by
the moving center sensing rod E provided with two capacitor plates and the side
plates. In balance position the capacitances are of equal value. A level
variation will a difference in buoyancy of the displacer. The center sensing
rod will move in the direction of one of the side capacitor plates. This causes
a difference in value of these capacitances. By an electronic circuit this
change is detected and integrated. During the rotation of the servo motor the
cam driven transmitter continuously change the voltage pattern to a remote
indicator of which the receiver motor drives a counter indicating level
variation.
.
1.
What is ;’cold junction compensation’?
‘Cold Junction Compensation’ is used in
temperature measurement by means of a thermocouple. This compensation is to
correct the error caused by the room (ambient) temperature. The mv produced by
a thermocouple is proportional to the temperature difference between its ‘hot’
and ‘ Cold’ junction. The cold junction is the ambient temperature (control
room). Without ‘cold junction compensation’ the temperature at the hot junction
will measure inaccurately.
2.
What is an RTD? How does it work?
RDT – Resistance Temperature Detector
RTD is a positive temperature coefficient
resistance which provides a linear rise in its resistance verses the rise in
temperature.
3.
What is the commonly used RTD in P.D.O.?
Pt 100 platinum resistance 100 Ohms
12. What is ‘pt 100’? What is the resistance
(in ohms) of a pt100 at 0 Deg C?
Pt 100 Platinum resistance which offers 100
ohms at 0 Deg C.
13.
What are the advantages and disadvantages of an RTD over
a thermocouple?
RTDs are accurate at a lower range of
temperature measurement such as -200 Deg C to +200 Deg C.
RTDs are expensive compared to Thermocouples,
also there are limitations in measuring at the higher range of temperature.
14.
Why does RTD measurement use 3 wires for a field signal
connection?
3 wire system is used in temperature
measurement by an RTD to compensate the line resistance. Three wire system provides a Wheatstone Bridge in the
measuring instrument.
15.
What is a ‘bi metallic’ temperature instrument? How does
it work? Give an example?
In a bi metallic temperature instrument, two
metals with different temperature coefficient of expansion are attached
together. Due to its expansion characteristics the higher coefficient of
expansion responds more than the lower one and a twist on the element is
formed. This principle is ;used in designing temperature measuring instrument
for the lower range.
Other than the filled type instruments, most
of the temperature indicators are of the bimetallic types.
16.
What is a ‘filled type’ temperature instrument? Give an
example?
Principle of change in the liquid expansion
with the change in temperature is used in designing the ‘filled type’
temperature measuring instrument.
For example : The glass thermometers, filled
type capillary temperature dial indicators and switches.
Filled type temperature measuring instruments
are used for lower range temperature measurement.
17.Name the
instruments used to measure temperature?
1.precision glass thermometers
2.RTD with a resistance measuring instrument
in terms of Deg C or Deg F.
3.Thermocouple with a mv measuring instrument
in terms of Deg C or Deg F.
5. FIRE & GAS SYSTEM
Fire
detection system
1.What is a
‘fire triangle’?
A fire triangle
represents the three elements, which causes a fire in a combustible mixture.
The three elements are fuel, air and ignition.
2.What is
importance or a ‘hood’ on a gas turbine?
During a gas
turbine normal running condition, a hood provides:
·
prevention
of turbine high dB noise to outside areas
·
keeps
the gas turbine clean from external dust
·
provides
a draft for the gas leak to the exhaust through the hood fan
During a gas
turbine shutdown condition a hood provides:
·
cooling
the turbine body by way of the hood fan
·
During
a fire shutdown it facilitates to put out the fire by confining the fire
extinguishers on the gas turbine.
·
3.Why do we use
‘Halon’ as a fire extinguisher in a gas turbine hood?
Halon is stored in
liquid form in a cylinder. When it is released in a hood during the occurrence
of the fire, it discharges the halon in gas from. Halon does not act or react
on electrical components.
4.What is the expansion from of B.C.F?
The expansion form
of B.F.C is Bromo chloro floride.
5.In a Ruston
gas turbine, why are there two Halon cylinders in each bank? How do they
function?
The first bottle is the ‘first shot’ and the second bottle is ‘Extended
shot’. The first bottle discharges the Galon into the hood through a 1’’ pipe
in approximately 15 seconds, where as the extended bottle discharges the Halon
through a ¼’’ pipe for another ½’’hour period to maintain the inert atmosphere.
6.How many UV detectors are installed in Solar, Ruston
TB-5000 and Ruston
TA-1750 gas turbine hoods?
Solar gas turbine hood: 4 UV detectors
Ruston TA-1750 hood: 4 UV detectors
Ruston TB 5000 hood: 12UV detectors
7.What is voting Logic of UV detectors in Soar and Ruston gas turbines?
There are two voting logics, they are: 1 out of 4 UV s and 2 out of 4 UV
s
8.What happens when UV detectors detect a fire?
1out of 4 UV s: creates annunciation, audible alarm on the control panel
and siren in the field (refer to the station drawings for the exact function
and operations).
2 out of 4 UV s: creates annunciation, audible alarm on the control
panel, a siren in the field, shutdown of the turbine and release of the fire
extinguisher (refer to the station drawings for the exact function and
operations).
9.How much is the time delay between fire sensing by a UV detector and
confirming with an alarm?
Generally it is set for 4 secs. The UV detectors initiates a fire alarm
only when the UV is detecting the fire continuously for 4 secs (refer to the
station drawings for the exact settings parameters).
10.What are the Halon manual release facilities available on a Solar gas turbine?
Auto Halon cylinder can be discharged manually from the fire & gas
control panel in the control room. On instiating the manual release:
·
The unit
shuts down.
·
The
auto Helon cylinder gets discharged in the hood.
·
Auto
halon discharge confirmation and the auto Halon cylinder pressure low alarm
appears on the control panel.
·
Audible
alarm in the control room and siren in the field occurs.
Manual Halon cylinder can discharged manually from the field through a
pall string. On initiating the manual release:
·
The
unit shutsdown.
·
The
manual Halon cylinder gets discharged in the hood.
·
Manual
Halon discharge confirmation and the manual Halon cylinder pressure low alarm
appears on the control panel.
·
Audible
alarm in the control room and siren in the field occurs.
11.Why are there ‘heat switches’ in side the hood, when there are UV
detectors?
Heat switches are a mechanical type and they are considered to be a
positive type of fire detection system. The heat switch settings are much
higher than the hood temperature.
12.What maybe the reasons if a unit shutdown on false heat detected
alarm?
The reason could be:
·
Hood
ventilation fan has failed or stopped.
·
Major
hot gas leak inside the hood.
13.What happens when a heat switch actuates?
On detection of heat, the heat switch initiates the following.
Annunciation, audible alarm on the control panel, a siren in the field,
shut down of the turbine and release of fire extinguisher (refer to the station
drawings for the exact settings parameters).
14.Wxplain the operating principle of a gas monitor?
Gas monitor measures the imbalance in the current loop caused by its
active and in-active filaments in the presence of a combustible gas.
15.Explain the calibration procedure of a gas monitor system.
·
Gas
sensor loop current or voltage at the sensor head is set as per the
manufacturer’s recommended value.
·
Gas
monitor zero is adjusted to the instrument air.
·
A test
gas with a known quantity of combustion gas (Generally methane 2% by volume) is
fed to the sensor and the span is adjusted to read 40% on the monitor scale
(refer to the station drawings for the exact settings parameters).
·
The
calibration procedure is repeated until the zero and span reads correctly.
16.What are the alarm and shut down setting on a gas monitor?
Generally on the gas monitor, the alarm is set at 20%rising and the shut
down is set at 60%rising (refer to the station drawing for the exact settings parameters)
17.What has to be done prior to entering a gas turbine hood?
Prior to entering a gas turbine, turn on the ‘ventilation defeat
switch’.
18.What is a ventilation defeat switch?
On activating, the ventilation defeat switch inhibits the release of
Halon in the auto mode and also the unit shutdown on ‘ventilation failure’.
19.How much is the delay between fire detection and Halon release? Why
is the time delay required?
A time delay of 15 secs. Is set between the detection of fire and the
initiation of Halon release. This is to achieve effective fire extinguishing by
allowing the hood fan to run-down to zero speed and the bleed valves to release
the compressed air.
20.What does a hood ventilation fan do when a fire is detected? And why?
On detection of fire, the fire system initiates the hood fan shutdown.
This is to minimize the presence of air the hood for releasing the
extinguisher.
Gas detection system
1.What is a ‘combustible gas mixture’?
Combustible gas mixture is the proper ration of fuel and air within the
ignition limits.
2.Explain the operating principle of a gas detection system.
Gas detection system measures the imbalance in the current loop caused
by its active and in-active filaments in the presence of a combustible gas.
3.Why is gas detection as important as fire detection?
Presence of combustible gas Is the first stage of a possible occurrence
of a fire.
4.What are the components used in a gas detection loop?
A gas sensor, 3 wire system and a gas monitor.
5.What is a ‘flame arrester’ in a gas sensor? What does it do?
Flame arrester is used on the sensors head. The sensors are continuously
powered to burn the combustible gas for measuring. The flame arrester prevents
the occurrence of fire outside the sensor.
6.What are the name of the filaments inside the gas sensor?
There are two filaments. They are called ‘active’ and ‘non-active’
filaments.
7.How much approximately is the resistance of a gas sensors filament?
From manufacturer to manufacture the sensor resistance varies. It is
approximately 3 ohms on each of the filament (refer to the station drawings for
the exact parameters).
8.How to check a gas monitor without test gas?
Some gas monitors are provided with a self test facility to check its
alarm and shutdown functions(levels).
9.Why is a three wire system used on the sensors?
The three-wire system is used to compensate the line resistance. The
active and non-active filament loop form a whetstone bridge in the gas monitor.
The amount of unbalance in the current loop during the occurrence of
combustible gas is detected as the percentage of combustible gas.
10.What are the types (manufacturer) of gas monitor systems used in
P.D.O.?
The commonly use gas monitor systems are:
1.
General
monitor
2.
Seiger
monitor
3.
Detection
instruments
11.What is test gas used to calibrate a gas monitor?
Methane gas is used for calibrating a gas monitor.
12.How much is the gas concentration in the test gas cylinder?
Generally methane 2% by volume is used as a test gas to calibrate a gas
monitor.
13.What is the important factor set on a gas monitor loop?
Depending on the manufacturer, either it is the loop current or the
voltage at the sensor head which has to be set precisely for better accuracy
and sensitivity.
14.why is it that Halon is not released on a high gas shutdown alarm in
a gas turbine hood?
Halon (fire extinguishing agent) is not released on detection of
combustible gas because the presence of gas is not a fire.
15.Why are there gas sensors on the combustion air intake of Ruston TB-5000gas
turbines?
The presence of combustible gas in the air intake may result in the
failure of stopping the gas turbine during a shutdown, due to the fuel and air
mixture continuously available to the turbine.
Smoke detection system
1.What are the types of ‘smoke detectors’ in use in P.D.O.?
Two types of smoke detectors are used in PDO. They are:
1.
Optical
type, 2. Radioactive type, 3. Ionisation type.
2.Explain the operating principle of the different types of smoke
detectors.
Optical type: when
the smoke cuts the optical detection path, and unbalance in the detector
current loop raises a smoke alarm. These type detectors are sensitive to larger
particles of combustion 0.5 to 10 microns. It is suitable for applications
where thermal turbulence can be expected.
Radio active type:
when the smoke is detected by the radio active substance, the unbalance in the
detector current loop raises a smoke alarm. This type of detectors are
sensitive to air movement greater than 5m/sec. It is normally installed with
the photoelectric (optical type) type smoke detectors.
3.Why is a smok4e detector as important as a ‘heat switch’ and ‘UV
detector’?
Smoke detectors are used in confined areas such as control rooms and
electrical switch rooms.
4.Where are smoke detectors generally installed in P.D.O.?
Smoke detectors are used in confined areas such as control rooms and
electrical switch rooms.
5.What happens when a smoke detector detects smoke?
When a smoke detector detects smoke, the detector loop current changes.
The unbalance in the loop current is detected as smoke.
Once in every 10 secs the L.E.D on each smoke detector blinks as self
loop check. On detection of smoke, this L.E.D remains continuous on (bright).
6.name the manufacturers of smoke detection systems.
Name of the smoke detector manufactures are Firecheta, Cloride Gent…etc.
7.why are periodic checks and maintenance required on smoke detection
system?
Particularly on the optical type smoke detectors, the collection of dust
in the sensor path may result in a false alarm. Also the periodic maintenance
ensures the reliability of the detection system.
8.What are the common problems faced when maintaining a smoke detection
system?
·
False
alarm due to accumulation of dust, floor cleaning, water vapour
·
Loose
connection results in the fluctuation of the detector loop current.
Improper base
plug-in connection.
1.What is a ‘combustible gas mixture’?
Combustible gas mixture is the proper ration of fuel and air within the
ignition limits.
2.Explain the operating principle of a gas detection
system.
Gas detection system measures the imbalance in the current loop caused
by its active and in-active filaments in the presence of a combustible gas.
3.Why is gas detection as important as fire detection?
Presence of combustible gas Is the first stage of a possible occurrence
of a fire.
4.What are the components used in a gas detection loop?
A gas sensor, 3 wire system and a gas monitor.
5.What is a ‘flame arrester’ in a gas sensor? What does
it do?
Flame arrester is used on the sensors head. The sensors are continuously
powered to burn the combustible gas for measuring. The flame arrester prevents
the occurrence of fire outside the sensor.
6.What are the name of the filaments inside the gas
sensor?
There are two filaments. They are called ‘active’ and ‘non-active’
filaments.
7.How much approximately is the resistance of a gas
sensors filament?
From manufacturer to manufacture the sensor resistance varies. It is
approximately 3 ohms on each of the filament (refer to the station drawings for
the exact parameters).
8.How to check a gas monitor without test gas?
Some gas monitors are provided with a self test facility to check its
alarm and shutdown functions(levels).
9.Why is a three wire system used on the sensors?
The three-wire system is used to compensate the line resistance. The
active and non-active filament loop form a whetstone bridge in the gas monitor.
The amount of unbalance in the current loop during the occurrence of
combustible gas is detected as the percentage of combustible gas.
10.What are the types (manufacturer) of gas monitor
systems used in P.D.O.?
The commonly use gas monitor systems are:
1.
General
monitor
2.
Seiger
monitor
3.
Detection
instruments
11.What is test gas used to calibrate a gas monitor?
Methane gas is used for calibrating a gas monitor.
12.How much is the gas concentration in the test gas
cylinder?
Generally methane 2% by volume is used as a test gas to calibrate a gas
monitor.
13.What is the important factor set on a gas monitor loop?
Depending on the manufacturer, either it is the loop current or the
voltage at the sensor head which has to be set precisely for better accuracy
and sensitivity.
14.why is it that Halon is not released on a high gas
shutdown alarm in a gas turbine hood?
Halon (fire extinguishing agent) is not released on detection of
combustible gas because the presence of gas is not a fire.
15.Why are there gas sensors on the combustion air intake
of Ruston
TB-5000gas turbines?
The presence of combustible gas in the air intake may
result in the failure of stopping the gas turbine during a shutdown, due to the
fuel and air mixture continuously available to the turbine.
16.Explain the calibration procedure of a gas monitor
system.
Gas sensor loop current or voltage at the sensor head is set as per the
manufacturer’s recommended value.
Gas monitor zero is adjusted to the instrument air.
A test gas with a known quantity of combustion gas (Generally methane 2%
by volume) is fed to the sensor and the span is adjusted to read 40% on the
monitor scale.The calibration procedure is repeated until the zero and span
reads correctly.
17.What are the alarm and shut down setting on a gas
monitor?
Generally on the gas monitor, the alarm is set at 20%rising and the shut
down is set at 60%rising (refer to the station drawing for the exact settings
parameters)
1.What are the types of ‘smoke detectors’ in use in P.D.O.?
Two types of smoke detectors are used in PDO. They are:
1.
Optical
type, 2. Radioactive type, 3. Ionisation type.
2.Explain the operating principle of the different types
of smoke detectors.
Optical type: when
the smoke cuts the optical detection path, and unbalance in the detector
current loop raises a smoke alarm. These type detectors are sensitive to larger
particles of combustion 0.5 to 10 microns. It is suitable for applications
where thermal turbulence can be expected.
Radio active type:
when the smoke is detected by the radio active substance, the unbalance in the
detector current loop raises a smoke alarm. This type of detectors are
sensitive to air movement greater than 5m/sec. It is normally installed with
the photoelectric (optical type) type smoke detectors.
3.Why is a smok detector as important as a ‘heat switch’
and ‘UV detector’?
Smoke detectors are used in confined areas such as control rooms and
electrical switch rooms.
4.Where are smoke detectors generally installed in
P.D.O.?
Smoke detectors are used in confined areas such as control rooms and
electrical switch rooms.
6.name the manufacturers of smoke detection systems.
Name of the smoke detector manufactures are Firecheta, Cloride Gent…etc.
7.why are periodic checks and maintenance required on
smoke detection system?
Particularly on the optical type smoke detectors, the collection of dust
in the sensor path may result in a false alarm. Also the periodic maintenance
ensures the reliability of the detection system.
8.What are the common problems faced when maintaining a
smoke detection system?
False alarm due to accumulation of dust, floor cleaning, water vapour
Loose connection results in the fluctuation of the detector loop
current.
Improper base plug-in connection.
Great! The info about level measurement was great and helpful too.. Keep Posting!!!
ReplyDeleteLevel Transmitters
Tank Calibration Services from EIWAA Group help avoid systematic storage losses on shore or at sea. Find out more.
ReplyDeleteBasically cavitation is a phenomenon that occurs during flow of a liquid in a closed channel such as a pipe. Normally when a liquid flows over any distance, there a drop in pressure – which is essential to maintain the flow.At some point this pressure drops much below even atmospheric pressure – particularly in case of water flowing in a pipe. When this happens water in the pipeline tends to evaporate. This water vapor forms bubbles within water body and these bubbles keep on popping up as well as collapsing. Any solid surface facilitates this process. as such surface of the pipe suffers maximum variation in pressure.
ReplyDeleteI am really impressed with your blog article, such great & useful knowledge you mentioned here. Your post is very informative. I have read all your posts and all are very informative. Thanks for sharing and keep it up like this.
ReplyDeleteSupply and installation fire system
Thanks for sharing this informative post. It's very helpful. Keep it up!
ReplyDeleteVacuum Gauge Tester | Laboratory Testing Equipment & Instruments
Such a nice blog, keep it up. also thanks to share this useful information
ReplyDeleteThanks for this post...Eiwaa Group specialise in analysing Storage Tank Survey and provide highly accurate strapping tables, shell deformation analysis, surveys for tank bottom and foundation settlement, shell roundness, structural column verticality, and more.
ReplyDeleteEddy Current Testing
Tank Calibration Services
Pipe Freezing Services
Calibration Services
ReplyDeleteGreat post! This is a useful post. This article is understandable and contains a lot of useful information. Thanks for the resume!
Thank you for the article. Comments are the best way to contact users and learn about their experiences.